Keep original data, confirm extraction and separate measurement from interpretation.

The report’s scope

ICP-OES and ICP-MS are elemental analysis techniques. A commercial report can combine several methods; check which produced each result. Do not attribute pH, salinity, alkalinity or organic-compound measurements to plasma analysis merely because they appear alongside elements. ReefFrag interprets existing reports and does not process physical samples.

Confirm transcription first

Upload the original and record laboratory, sample, date and tank. Check analyte, value, unit, reference and detection limit. Leave unreadable values unresolved. Zero, not detected, below limit and not measured are different states. Confirm the table before requesting an explanation.

N-DOC explained correctly

TRITON N-DOC reports TIC, total inorganic carbon; TOC, total organic carbon; and TNb, total bound nitrogen. It does not mean Non-Dissolved Organic Carbon and is not exclusively a detritus measurement. It complements elemental information using other methods. Do not convert TNb directly to nitrate or equate TIC alone with total-alkalinity titration. Check the report and method.

References and priorities

A laboratory reference differs from your configured tank target. Evaluate deviations using uncertainty, trends and livestock observations. A single result does not prove the cause of tissue loss. Identify missing information, potential contamination and an independent confirmation method.

Compare and act

Compare reports only after checking units, methods and sampling date. Elemental phosphorus is not automatically orthophosphate. Prioritize error confirmation and stability before chasing every coloured flag. Log actions and follow-up; dosing requires verified volume, product and concentration.

A cautious reading example

One report shows copper below a limit and a later report gives a number. Before declaring new contamination, compare methods, limits, units and sampling. Check laboratory or container changes. A below-limit value is not known zero and cannot establish an exact increase from zero. For confirmed deviations, investigate inputs and obtain appropriate confirmation before attributing symptoms to one source.

How an explanation should read

A useful interpretation starts with confirmed information and separates priorities from hypotheses. State what users must verify, missing context and observations or tests that could change the conclusion. Preserve laboratory references rather than silently replacing them with tank targets. Declare manual extraction and leave unreadable values unresolved. AI must not complete a plausible-looking table or turn low elements into automatic supplementation.

A comparison that retains evidence

Keep originals, sampling dates, confirmed tables and contemporaneous explanations. When correcting a unit, retain the previous version and reason. Do not plot one continuous trend across changed analyte definitions. Total element, dissolved fraction and specific compound may need separate series. Downloaded reports should show missing data and pending confirmation so summaries do not appear more conclusive than the evidence.